50
DD1010 FLYNET
FIELDS
Fields
Defines the number of weighing ranges into which the nominal
capacity of the instrument is divided.
It is possible to select:
1:
the indicator has a single weighing range for the entire range
of the scale capacity.
2MD:
the indicator has two weighing ranges with MULTI-
DIVISION mode
3
.
2ME:
the indicator has two weighing ranges with MULTI-
EXTENSION mode
4
.
3MD:
the indicator has three weighing ranges with MULTI-
DIVISION mode.
3ME:
the indicator has three weighing ranges with MULTI-
EXTENSION mode.
Range1 Capacity
Value of the capacity corresponding to the first partial weighing
range.
Range2 Capacity
Value of the capacity corresponding to the second partial
weighing range. (in case of 3 ME or 3 MD)
W1 extension return delay (no.
of conversions)
Number of samplings of delay on return to extension W1 for
multi-extension scales.
Manual range management
Possibility of selecting the ranges manually.
SUM PARAMETERS
A, B, C...
Indicates which of the scales present are used to create the
sum scale
3
multi-division = means that to use again the smallest division it is necessary to return within the first weighing range
4
multi-extension = means that to use again the division of W1 it is mandatory to pass through 0