DD1010 FLYNET
DANGER
It is mandatory to use the Electrostatic Discharge Sensitive (EDS) cuff before performing any
operation inside the indicator to avoid damaging the circuit boards or parts of the indicator.
For connections and technical information, refer to the diagrams and instructions in the manuals
supplied.
Key:
1. Stamped self-adhesive label sealing the access hole to the calibration switch
2. Calibration switch
Referring to the previous figures, proceed as follows:
Switch the indicator on.
Remove the self-adhesive label (ref. 1) which seals the access hole to the calibration switch.
Using a non-metal tool, press the calibration switch (ref. 2).
ATTENTION
Removing the metrological stamps makes it impossible to use the weighing indicator for
commercial transactions towards third parties. For their restoration contact the After Sales
Service
of
the
Service
Network
http://www.coopbilanciai.it/it/home/rete-vendita-e-
selecting your country
).
9.2
Version with cable glands
Figure 9.2-1 - Calibration button - Stainless steel indicator version
Key:
1. Indicator opening protection seal
2. Stamped self-adhesive label sealing the protective screw of the calibration switch
3. Calibration switch
Referring to the previous figure, proceed as follows:
Switch the indicator on.
2
1
3