DD1010 FLYNET
DANGER
Check that:
The power supply line of the indicator has voltage and frequency as indicated on the
stamped plate on the indicator bottom;
The socket to which the indicator is connected is provided with grounding pin;
Warning and danger plates are present on the hood of the instrument;
If this is not the case, inform the maintenance personnel or our Service Department
directly.
Scale cable wiring
A pre-wired cable is normally supplied for connecting the indicator to the weighing platform.
Key
1. 9/15 pin male D-sub connector
2. Nut
3. Indicator back wall
4. Cable gland
5. Shielded shell (cap) with stamping screws
Figure 3.2-4 - Load cell connection
Insert the 9/15 pin connector in the scale input hole
Lock the cable gland with the nut.
Refit the shell with the drilled stamping screws.
Insert the connector on the bottom of the CPU housing.
The connection can be different depending on the type of cell that is fitted on the weighing platform:
analogue or digital. In case of a duplex scale, connect scale A to connector JBIL1, and scale B to
JBIL2.
WARNING
The cable shield must always be connected to the metal protection cap of the 9/15 pin connector.
Do not insert the connection cable to the weighing platform into routes with cables directed to
power drives.