LB 571 BTwave
8 Error Codes and Alarms
68847BA2 Rev.00, 10/2020
41
8
Error Codes and Alarms
8.1
Watchdog
It is possible to use on-chip hardware watchdog timer to recover software dead-
locks. Watchdog timer can be enabled in
device_parameters #99
(Watchdog_enable)
. Watchdog timer timeout is fixed to 5 seconds.
8.2
Error codes
There are following general error codes defined. Errors are inserted to the error
code word bit by bit. Each error corresponds one bit so totally 16 errors can be
defined. Error code is read by UI in diagnostics tab.
Errors are cleared automatically if they disappear. Watchdog error is only cleared
by powering down the resonator.
Error types in error code can’t be masked. All error types are always active. In other
hand errors which generate alarm can be defined.
Error Codes
Error code
Value
Description
SENSOR_RESONANCE_LEVEL_ERR 0x0001 Resonance level under threshold value. Threshold
defined using device parameters #92
SENSOR_TEMP_MATERIAL_ERR
0x0002 Unable to read material temperature sensor
SENSOR_TEMP_CERAME_ERR
0x0004 Unable to read cerame temperature sensor
SENSOR_ANALOG_OUT_1_ERR
0x0008 Unable to write to Iout channel 1
SENSOR_ANALOG_OUT_2_ERR
0x0010 Unable to write to Iout channel 2
SENSOR_SD_CARD_WRITE_ERR
0x0020 Unable to write to the SD-card.
SENSOR_PLL_CONFIG_ERR
0x0040 Phase locked loop was not configured correctly at
startup
SENSOR_RTC_ERR
0x0080 Real-time clock not setup or has been resetted
SENSOR_WD_RESET_ERR
0x0100 Indicates that last reset was caused by watchdog.
Is cleared by power on reset only.
SENSOR_HEATING_CTRL_ERR
0x0200 Indicates that FET which controls ceramic heating
is not functional
SENSOR_SD_CARD_INIT_ERR
0x0400 SD-card initialization is not succeeded. There is
problem to mount file system in startup or after
re-inserting SD-card. Writes to SD-card can’t be
made.
Summary of Contents for BTwave LB 571-02
Page 1: ...BTwave LB 571 02 Operating Manual 68847BA2 Rev No 00 01 2020 ...
Page 2: ......
Page 4: ......
Page 24: ...5 Commissioning LB 571 BTwave 20 68847BA2 Rev 00 10 2020 ...
Page 44: ...7 Diagnostics LB 571 BTwave 40 68847BA2 Rev 00 10 2020 ...
Page 48: ...8 Error Codes and Alarms LB 571 BTwave 44 68847BA2 Rev 00 10 2020 ...
Page 50: ...9 Technical Information LB 571 BTwave 46 68847BA2 Rev 00 10 2020 ...