6 Start up and Calibration
LB 571 BTwave
28
68847BA2 Rev.00, 10/2020
6.7
Calibration
In order to have device measuring moisture content, it must be calibrated using
the material it will be measuring. Refence samples should be taken covering the
whole moisture range used at the plant.
6.7.1
Sampling
Before starting with sampling, please enssure that the SD-card is inserted in the
device.
To start taking samples, go to
Calibration
tab shown on below figure and follow
the steps shown below:
Figure 18 Taking samples
1.
Click New sample…
2.
Below Moisture measurement view opens
Figure 19 start sampling
3.
Click
Start
(button toggles to
Stop
) and start collecting a sample for reference
determination.
4.
When finished sampling, click
Stop
button.
5.
The above steps 3 and 4 can be repeated.
6.
Finally click
Close
button.