LB 571 BTwave
7 Diagnostics
68847BA2 Rev.00, 10/2020
37
7.3
Writing and reading product parameters
Recipe parameters are saved to internal FLASH memory after Activate Parameters
is selected from UI. In addition recipe parameters are saved to SD-card parameters
–folder. Parameters are never copied from SD-card to internal FLASH.
Selecting Get Active Parameters loads recipe parameters from sensor FLASH
memory to UI.
Figure 32 Recipe/Product parameters
7.4
Inserting and removing SD-card on the fly
It is possible to remove and insert SD-card to the device on the fly. If device is writ-
ing data to the SD-card when card is removed MCU SW may get stuck.
It is possible to use a watchdog timer d
evice parameters #99 (Watchdog_enable)
resetting the MCU after 5 seconds. Current output is not affected during reset and
all parameters are recovered from internal FLASH memory to continue measure-
ment as before system reset. Device is also capable of doing measurements without
SD-card inserted.