Save reference values to samples
UI must have reference values for samples used in calculating moisture measure-
ment parameters for the device. This is also done under Calibration tab and as
shown below:
Figure 21 Calibration view with samples
1.
Double-click on a sample shown above and Calibration File view below
opens, where you can give additional information for the sample.
Figure 22 Save additional informationtion to sample file
2.
Type reference value on
Lab %
box
3.
Type used product on
ProductID
box. This is by default set to the product,
which has been active when sample was taken.
4.
Type comments on
Additional Information
box
5.
Click
Save
and view closes
It is also possible to remove samples from SD-card and PC by clicking
Re-
move file.