6 Start up and Calibration
LB 571 BTwave
26
68847BA2 Rev.00, 10/2020
13# AO2_scale_max (°C)
: Analog output channel 2 maximum material temperature
(°C) value. Default 100 °C.
25# Product_choice
:
Product ID selection automatically from binary inputs or
manually set by UI. 0=Auto (Selection made by binary
inputs), 1=MAN (See Param #26). Default 0.
26# Product_nr:
If Param 25# set to manual selection, this can be used for
changing product ID number (0-3) from UI: e.g. 0=material
1, 1=material 2, etc.. (See Param 25#).
33# Output_filtering
:
Output filtering of moisture reading in seconds (1-300
seconds). Default 20 s.
90# Temp_reg_hi (°C)
:
High limit for ceramic temperature control (°C). Enter
max. product temperature. Default 40°C
91# Temp_reg_low (°C)
: Low limit for ceramic temperature control (°C). Enter
max. product temperature-0.1°C. Default 39.9 °C.
98# SD_card_detect
:
Shows if SD-card is inserted or not. 0=no SD-card inserted,
1=SD-card inserted. This is a read-only parameter on the
device, but if set to 0, sample file will only be saved on the
PC and can’t be found on the device afterwards.
Setting ceramics temperature
For compensation and stability issues temperature of ceramics should be set:
Parameter 90# (see figure 9) = max. product temperature
Parameter 91# (see figure 9) = max. product temperature -0.1°C
Summary of Contents for BTwave LB 571-02
Page 1: ...BTwave LB 571 02 Operating Manual 68847BA2 Rev No 00 01 2020 ...
Page 2: ......
Page 4: ......
Page 24: ...5 Commissioning LB 571 BTwave 20 68847BA2 Rev 00 10 2020 ...
Page 44: ...7 Diagnostics LB 571 BTwave 40 68847BA2 Rev 00 10 2020 ...
Page 48: ...8 Error Codes and Alarms LB 571 BTwave 44 68847BA2 Rev 00 10 2020 ...
Page 50: ...9 Technical Information LB 571 BTwave 46 68847BA2 Rev 00 10 2020 ...