6 Start up and Calibration
LB 571 BTwave
32
68847BA2 Rev.00, 10/2020
The moisture calculation parameters are optimized by the UI for best correlation
between the measured signals and the Lab-% values. The calculation parameters are
shown in the quadratic box on the right.
The moisture reading is based on multivariable density compensated measurement.
The calibration may be improved by using the below calculation methods:
Temperature
:
If material temperature range for the collected samples is big,
then temperature compensation should be used by ticking the
Tempereture
check box.
Polyfit
:
If this is ticked parameter Coeff A will affect the calculation (value
different than 0).
Should not be used until the whole moisture range is covered by
the samples.
If material is quite homogenous and the whole moisture range is
covered by the samples, the parameter optimization may be used
without the
Polyfit
.
The statistics for the calibration is shown in the bottom right corner. These are as
follows:
error:
RMSE (Root Mean Square Error) between the difference of Measured-%
and Lab-% of the samples.
steyx:
Comparison between Measured-% and Lab-% of the samples using excel
function STEYX.
cor:
Correlation between Measured-% and Lab-% of the samples using excel
function CORREL.
5.
When parameter optimization is done, click
Move Parameters to Pending
List …
button marked with the arrow in the autocalibrator
6.
Below view opens with the active material selected
Figure 25 Save parameters to pending list
7.
Select calibrated material, write a comment and click
OK
.
8.
UI opens
Parameters History
tab under the
Parameters
tab. See below fig-
ure.
Summary of Contents for BTwave LB 571-02
Page 1: ...BTwave LB 571 02 Operating Manual 68847BA2 Rev No 00 01 2020 ...
Page 2: ......
Page 4: ......
Page 24: ...5 Commissioning LB 571 BTwave 20 68847BA2 Rev 00 10 2020 ...
Page 44: ...7 Diagnostics LB 571 BTwave 40 68847BA2 Rev 00 10 2020 ...
Page 48: ...8 Error Codes and Alarms LB 571 BTwave 44 68847BA2 Rev 00 10 2020 ...
Page 50: ...9 Technical Information LB 571 BTwave 46 68847BA2 Rev 00 10 2020 ...