36
2. When the value of the calibration weight begins to flash, load the calibration weight on the
plate.
3. Wait for the calibrated weight to be displayed and the stability symbol to turn on, and then
remove the weight from the plate.
4. When
“0.000”
is shown on the display, press the
button in a prolonged manner
until the beeping stops.
The acquisition and automatic storage of the internal weight will
now begin. During the acquisition cycle, the display will show the following flashing
writing:
5. Once the internal calibration has been stored, the balance will return to the normal
weighing condition.
6. Now re-enter the calibration menu as described in paragraph 9.1.12 and set the desired
internal, automatic, or external calibration modality.
ATTENTION : this procedure must be effected only using E1 class reference
masses.
Mem. Tech
01-06-10 13:15:24
200
g
01-06-10 13:15:24
-0-