Weighing
After connecting the balance to power supply, an autodiagnosis of electronic circuits is
automatically effected, ending with stand by indication
Balance Warm-up: Wait
8 hours from switch on for warm up
It is suggested to never disconnect the balance from power socket and use
ON/OFF key to put the instrument in standby mode until end of work.
From
“STAND BY”
mode: to bring the balance back to working conditions, press
ON/OFF
key
.
It is recommended not to drop heavy objects on balance pan, to avoid damage of the
instrument.
Electronic balance effects mass measurings using gravity (g). Differences in geographical
areas e in altitude change gravity acceleration (g).
Therefore, to get precise measurings, balance has to be adjusted to environmental
conditions. This adjusment is accomplished through calibration function.
It is needed to calibrate the balance every time it is moved to another place.