NOTE: an error message will be displayed if there is some interference during the calibration process.
To interrupt the calibration process, press the ON/OFF button while the calibration weight indication
flashes.
8.2 Balances with internal calibration
In these models the calibration is carried out through an internal automatic system:
1. Press the
CAL
button with the plate empty.
The display will show the message
“CAL”
and the balance’s calibration will be carried out
automatically.
2. At the end of the calibration, the balance will return to normal weighing conditions.
If the calibration is not completed due to vibrations or drafts, the message
“CAL bUT”
will
be displayed. Press the CAL button again, and if the problem persists, select external
calibration and contact the supplier.
To modify the calibration mode in these models with internal calibration, see section 9.12.1