6 - Maintenance
Figure 16 - Air Filter Replacement
6.2.2 Temperature Verification
(QC)
A thermometer and special vial are provided for temperature verification (QC)
of instrument incubation. It is recommended that the accuracy of the thermom-
eter be verified against a calibrated lab thermometer to insure the validity of the
temperature verification.
The temperature control circuitry is designed to maintain the cabinet tempera-
ture at 35° C ±1.5° C. If your manual reading is within 1.5° C of the setpoint (35°
C), the controller and heaters are operating within their specifications.
The temperature vial is mounted in a bracket on the inside of the instrument
door (see Figure 2).
MA-0103-5