PROCEDURE
Now
vow
Specimen labeling:
Each vial should be labeled with the appropriate patient information:
• Patient Identification
• Laboratory Log Number
• Patient Location
• D a t e and Time of Collection
• Collector's Initials
• S i t e of Venipuncture
Fntering new vials:
• Press the 'Vial Entry' key.
• S c a n vial barcode by placing the vial in the alignment block in front of the scanner with the
barcode facing the scanner. Rotate the vial slightly in front of the scanner. The system
beeps once to indicate a successful scan.
NOTE: I f the vial barcode cannot be read, press the 'No Barcode' key. Select the vial
media by using the Up/Down arrow keys. Press the 'Ok' key to confirm media type.
• I n s e r t the vial into position indicated on the LCD display.
• I f default protocol is acceptable, press `Ok' key.
NOTE: To modify the protocol length for a particular vial, select the 'Change Protocol'
key. Select the desired protocol length using the Up/Down arrow keys. Press 'Ok' to
confirm protocol length.
• V e r i f y that all information is correct and that vial has been inserted into the indicated station.
Press 'Ok' key to confirm.
• Repeat the above steps for each new vial. Then Press the 'Exit' key.
Positive Vials:
A. Notification of the presence of presumptive positive vials
• A n audible alarm sounds if configured to a volume > O.
• T h e New Positive Indicator on the front of the instrument flashes red.
• T h e 'Remove Positives' key appears in the LCD display when the door is open.
• O n the main status display. that station with the positive vial is displayed as a filled circle
with a plus
sign in
it, and the positive total in the summary region reflects the number of
positives in the instrument.
B. Removal of positive vials
• Press the 'Silence Alarm' key to acknowledge the alarm.
• O p e n instrument door and press the `Remove Positives' key.
• L C D display identifies the positive vial by position and the barcode number is also
displayed.
• Remove vial from the station and scan the vial barcode. Repeat the above steps until all
positives are removed (three beeps indicates all positives have been removed).
• Press the 'Exit" key.
NOTE: A subculture and a Gram stain should be performed from each presumptive
positive vial.