Remove the vial from the station currently displayed on the vial entry
screen.Write this vial sequence number on the chart, as well as the
actual station. Now scan the current vial label.Write the station now
displayed in the"assigned station" column. Place this vial aside for sub-
culturing.
Go to the station now displayed. If there is a vial in this station, repeat
the steps in the previous paragraph. Continue to do so until the actual
station matches the assigned station, or until there is no vial in the
assigned station.When one of these conditions occurs, you have prob-
ably reached the end of the vial misplacements.
1 A l l vials involved in misplacement scenarios must be subcultured.
D W h e n the next test cycle or rotor scan occurs, any stations from which
vials were removed will go into error. In addition, the anonymous sta-
tion that began the problem is still anonymous. Go to the identify anony-
mous activity, and recall the anonymous vial that originally generated
this error. Select any media type, then press the "OK" soft key. Press
the "OK" soft key again at the confirmation prompt.Allow this station
to go into error too, then resolve all the station errors as described
above under error E 12 and in Section 4.7 after the misplaced vials have
been subcultured.
E34 U p d a t e error
POSSIBLE CAUSE(S)
During a software update, an error occurred.
CORRECTIVE ACTION(S)
Return to the "update software" operation and repeat the software
update.
E50 I n t e r n a l software error
POSSIBLE CAUSE(S)
System encountered a software error.
CORRECTIVE ACTION(S)
0 W r i t e data to disk and call Becton Dickinson.
•