Remove the vial from the station currently displayed on the vial entry
screen. Write this vial sequence number on the chart, as well as the
actual station. Now scan the current vial label.Write the station now
displayed in the "assigned station" column. Place this vial aside for sub-
culturing.
Go to the station now displayed. If there is a vial in this station, repeat
the steps in the previous paragraph. Continue to do so until the actual
station matches the assigned station, or until there is no vial in the
assigned station.When one of these conditions occurs, you have prob-
ably reached the end of the vial misplacements.
• A l l vials involved in misplacement scenarios must be subcultured.
• W h e n the next test cycle or rotor scan occurs,any stations from which
vials were removed will go into error. Resolve all the station errors as
described above under error E12 and in Section 4.7 after the mis-
placed vials have been subcultured.
E31 D i s k e t t e error
POSSIBLE CAUSE(S)
Floppy disk is not inserted.
• F l o p p y disk is not formatted.
• F l o p p y disk is write-protected.
Floppy disk is full.
CORRECTIVE ACTION(S)
I I n s e r t formatted floppy disk.
I M o v e write-protect tab toward center of floppy disk.
O U s e a new floppy disk.
£ 3 2 I n s t r u m e n t full
POSSIBLE CAUSE(S)
During Vial Entry or Resolve Station Errors, the system tried to allo-
cate a station but found that none are available.
CORRECTIVE ACTION(S)
I R e m o v e final negative vials if any exist. If not, vial should be tested
manually.