Distorted scatter parameters
Possible causes
Recommended solutions
Cytometer settings are
improperly adjusted
Optimize the scatter parameters. See the
BD FACSDiva™ Software Reference Manual
for
instructions.
Air bubble in the sheath filter
or flow cell
Purge the air from the filter. See
Removing air bubbles (page 29)
.
Flow cell is dirty
Flush the system. See
Air leak at sheath container
Ensure that the sheath container lid is tight and all connectors are secure.
Hypertonic buffers or fixative
Replace the buffers or fixative.
Excessive amount of debris in display
Possible causes
Recommended solutions
Threshold level is too low
Increase the threshold level.
Sheath filter is dirty
Replace the filter. See
Changing the sheath filter (page 43)
.
Flow cell is dirty
Flush the system. See
Dead cells or debris in the
sample
Examine the sample under a microscope.
Sample is contaminated
Re-stain the sample. Ensure that the tube is clean.
Stock sheath fluid is
contaminated
Rinse the sheath container with DI water, then fill the container with sheath fluid from
another (or new lot) bulk container.
High CV or poor QC results
Possible causes
Recommended solutions
Air bubble in sheath filter or
flow cell
l
Purge the filter. See
Removing air bubbles (page 29)
l
Prime the fluidics system. See
Preparing the fluidics (page 32)
Sample flow rate is set too
high
Set the sample flow rate lower.
Air leak at sheath container
Ensure that the sheath container lid is tight and all connectors are secure.
Flow cell is dirty
Flush the system. See
Waste tank is pressurized
Replace the waste container cap. See
Replacing the waste tank cap (page 42)
Poor sample preparation
Repeat sample preparation.
Sample was not diluted in the
same fluid as the sheath fluid
Dilute the sample in the same fluid as you are using for sheath.
Old or contaminated QC
particles
Make new QC samples and perform the quality control procedure again.
94
BD FACSymphony™ A1 Flow Cytometer User's Guide
Summary of Contents for FACSymphony A1 Flow
Page 1: ...FACSymphony A1 Flow Cytometer User s Guide 23 23437 01 2022 07 For Research Use Only ...
Page 6: ......
Page 10: ...This page intentionally left blank ...
Page 24: ...This page intentionally left blank ...
Page 36: ...This page intentionally left blank ...
Page 50: ...This page intentionally left blank ...
Page 87: ...More information l Running a performance check page 55 Chapter 7 Technical overview 87 ...
Page 88: ...This page intentionally left blank ...
Page 96: ...This page intentionally left blank ...