Changing the Bal seal
Introduction
This topic describes how to replace the Bal seal.
The sample injection tube Bal seal is a ring that forms a seal with the sample tube and ensures proper tube
pressurization.
When to change the Bal seal
Over time, the Bal seal becomes worn or cracked and requires replacement. Replacement is necessary if a
proper seal is not formed when a sample tube is installed on the SIP. Indications that a proper seal has not
formed include:
l
The tube will not stay on the SIP without the tube support arm.
l
When the tube is installed and RUN is pressed on the cytometer, the RUN button is orange (not green).
Cytometer hardware might be contaminated with biohazardous material. Wear suitable protective
clothing, eyewear, and gloves whenever cleaning the cytometer or replacing parts.