No
Definition
1
Violet (405 nm)
2
Blue (488 nm)
3
Yellow-Green (561 nm)
4
Red (637 nm)
About configuration maps
Filter and mirror arrangement
The filters are arranged in the detector array to steer progressively shorter wavelengths of light to the next
detector in the array. The longest wavelength should be in the A position and the shortest wavelength should
be in the last position used.
There should not be any empty slots for any laser being used. Always use a blank optic holder.
If a slot is filled with a filter or mirror, an identifying number appears in that position on the configuration map.
If a slot is filled with a blank optic holder, that position on the configuration map is unlabeled.
Chapter 9 Detector array configurations
99