If any parameters did not pass, see the
BD Cytometer Setup and Tracking Application Guide
for
troubleshooting information.
12. Select
File > Exit
to close the CS&T window and return to the BD FACSDiva™ interface.
The CST Mismatch dialog opens.
Click the
Details
button to verify which cytometer settings will be updated.
13. Click
Use CST Settings
.
By selecting Use CST Settings, the laser delay, area scaling, and other cytometer settings will be updated to
the latest settings from the performance check.
Chapter 5 Optimizing cytometer settings
57