If the cytometer is not set to the correct configuration:
a. Click
Select Configuration
in the
Setup Control
window.
b. Select the correct configuration from the list.
c. Click
Set Configuration
and then click
OK
.
4. Verify that the current configuration has a valid baseline defined.
If not, see the
BD Cytometer Setup and Tracking Application Guide
for more information on defining a
baseline.
5. Prepare the CS&T beads according to the technical data sheet provided with the beads or available on the
BD Biosciences website (
).
6. Install the bead tube onto the SIP.
7. In the
Setup Control
window, select
Check Performance
from the
Characterize
menu.
8. Click
Run.
9. Ensure that Fine Adjust is set to 250, press RUN, and LOW.
Plots appear under the Setup tab and the performance check is run. The performance check takes
approximately 5 minutes to complete.
10. Once the performance check is complete, click
View Report.
11. Verify that the cytometer performance passed.
In the Setup tab, the cytometer performance results should have a green checkbox displayed and the word
Passed
next to it.
56
BD FACSymphony™ A1 Flow Cytometer User's Guide
Summary of Contents for FACSymphony A1 Flow
Page 1: ...FACSymphony A1 Flow Cytometer User s Guide 23 23437 01 2022 07 For Research Use Only ...
Page 6: ......
Page 10: ...This page intentionally left blank ...
Page 24: ...This page intentionally left blank ...
Page 36: ...This page intentionally left blank ...
Page 50: ...This page intentionally left blank ...
Page 87: ...More information l Running a performance check page 55 Chapter 7 Technical overview 87 ...
Page 88: ...This page intentionally left blank ...
Page 96: ...This page intentionally left blank ...