20 / 46
51376-010 GB/2020-10
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
3.2.4
Miscellaneous functions and settings
Additional functions are available within a second menu level. These can be accessed by
simultaneously pressing and holding the "SET" key and arrow key. Switching between these
functions is accomplished with the "SET" key in the manner already described. The menu level
can only be exited by pressing the "START/STOP" key. In doing so, settings and function statuses
are carried over and executed.
probe check
off
scan frequency
off
contrast 22
operating mode
Ampl.
display Energy
batch mode
off
03.01 - 2016-08-085
off
Check the ultrasonic converter
Search resonance frequency
Set monitor contrast
Set type of ultrasound control
Check the ultrasonic converter [probe check]
This function is used for quick veri
fi
cation of whether the ultrasonic converter is recognised by
and can be correctly activated by the ultrasonic generator. After a change in probe or in operating
conditions it is possible, for example, that the ultrasonic generator cannot synchronise with the
ultrasonic converter and issues an error message, e.g. Error 011. The frequency is reset to the
base value and the function is restored if no device fault or other cause is present.
The set value is "o
ff
". In order to activate the function, set the parameter to "start" using the arrow
keys
, and then press the "START/STOP" key. See chapter 6.2.1 for additional information.
Search for resonance frequency [scan frequency]
This function is an expansion of the above-mentioned "probe check" function. In addition, a
frequency scan is conducted in order to determine the correct ultrasonic frequency for the
ultrasonic converter. Depending upon the probe used and the operating conditions, the frequency
can
fl
uctuate signi
fi
cantly in some instances. When switched o
ff
, the ultrasonic generator saves
the current frequency value and uses it as the start value when switched on again. If there are
signi
fi
cant deviations in the current frequency, this can lead to malfunctions.
Important:
After a successful scan while in continuous operation (pulsation = o
ff
, time > 30 s), start with an
amplitude setting of 50 % and check the operation. Next, set the desired value.
Setting the monitor contrast [contrast]
With this function, the monitor contrast can be adapted to the local light conditions. The contrast
values range between 1 and 68, and can be set with the arrow keys
. Smaller values result
in a darker image and larger values result in a lighter image. The default value is 22.