Prog 1* Ampl
10%
Time set
00:02:15
elapsed
00:00:00
Pulse 4.0s 4.0s
Stop at
0°C
---°C
Energy 0.000
kJ
4
Prog 1* Ampl
10%
Time set
00:02:15
elapsed
00:00:00
Pulse 5.1s 4.0s
Stop at
0°C
---°C
Energy 0.000
kJ
4
Prog 1* Ampl
10%
Time set
00:02:15
elapsed
00:00:00
Pulse 5.1s 4.0s
Stop at
0°C
---°C
Energy 0.000
kJ
4
Prog 1* Ampl
10%
Time set
00:02:15
elapsed
00:00:00
Pulsation
off
Stop at
0°C
---°C
Energy 0.000
kJ
Pulsation 4.0s
4.0s
18 / 46
51376-010 GB/2020-10
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
3.2.2
Pulsation [Pulsation]
/4
In addition to setting the pulse turn-on time t
E
and the pulse turn-o
ff
time t
A
, two additional
operating modes can also be set:
o
ff
no pulsation or continuous sound
by hand key
it is possible to pulse manually with the key on the ultrasonic converter.
Ultrasound operation will be active as long as the key on the ultrasonic converter
is pressed.
a) Setting the operating modes
SET
Prog 1* Ampl
10%
Time set
00:02:15
elapsed
00:00:00
Pulsation
off
Stop at
0°C
---°C
Energy 0.000
kJ
Pulsation
off
Prog 1* Ampl
10%
Time set
00:02:15
elapsed
00:00:00
Pulsation
off
Stop at
0°C
---°C
Energy 0.000
kJ
Pulsation by handkey
Prog 1* Ampl
10%
Time set
00:02:15
elapsed
00:00:00
Pulsation
off
Stop at
0°C
---°C
Energy 0.000
kJ
Pulsation 4.0s
4.0s
b) Setting the time intervals
The maximum setting value is 600.0 s. The minimum setting value will depend on the type of
device and ranges between 0.1 and 0.3 s.
Turn-on time t
E
Turn-o
ff
time t
A
START
STOP
back to stand-by mode
SET
proceed to next operating parameter
/4 The pulse period duration is the sum of the turn-on and turn-o
ff
time t = t
E
+ t
A