Dismounting
Assembly
Sickle spanner
HS 40/42
Spanner
here: Spanner SW 10
29 / 46
51376-010 GB/2020-10
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
4.3.1.2 Mounting and dismounting of titanium probes
On UW 100/200, titanium probes (TS ...) are mounted onto standard/booster horn SH 100 G/
SH 200 G. On UW 50 they are directly mounted
First, carefully wipe the mounting surfaces of the standard/booster horn and the probe
ensure clean mounting surfaces.
Screw on the probe by hand.
Lay the probe on an approx. 20 mm thick base so that it does not bend.
Firmly hold the standard/booster horn in place using sickle spanner HS 40/42, and tighten the
probe using the appropriate spanner, see illustration. It is recommended to the use the torque
wrench DMS 10.
Note: Observe the torque. - see product information
With the UW 50 use sickle spanner HS 25/28 - the probe is mounted directly onto the UW 50.
Dismount in the reverse order.