37 / 46
51376-010 GB/2020-10
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
Error
Possible cause
Remedial action
Signi
fi
cant heating in the
vicinity of the mounting
surfaces between the
ultrasonic converter and
standard/booster horn or
between the standard/
booster horn and probe?
Sonicating parts (standard/booster
horn and probe) not mounted
fi
rmly enough?
Are mounting surfaces of the
sonicating parts soiled?
- Dismount the respective parts, clean
the surfaces and
fi
rmly retighten, see
chapter 4.4.
If it is not possible to rectify the fault using these short instructions, please contact your local
specialist dealer or write us to the following address.
6.4
Repairs and service
If you identify errors or defects that cannot be recti
fi
ed, the homogeniser may not be used.
In such a case, please contact the supplier or the manufacturer in writing and use the enclosed
questionnaire:
BANDELIN electronic GmbH & Co. KG
Heinrichstrasse
3-4
12207
Berlin
Repairs/ Maintenance Department:
E-mail:
Phone: +49-(0)-30 – 768 80 – 13
Fax:
+49-(0)-30 – 76 88 02 00 13
In the case of returns, the general terms of delivery of BANDELIN electronic GmbH & Co. KG shall
apply.
In addition, the homogeniser must be cleaned and decontaminated (if necessary), see the
following chapter.
Caution
• Only allow repairs to be conducted by authorised skilled personnel!
• Before each repair, turn o
ff
the device and disconnect the plug from the mains!
•
Defective parts may only be replaced with original parts.
• The black housing of the ultrasonic converter and the aluminium cylinder (ultrasonic oscillating
system) must not be twisted toward each other. The ultrasonic oscillating system and its
electrical connections would be damaged as a result.