19 / 46
51376-010 GB/2020-10
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
3.2.3
Temperature monitoring [Temperature]
The temperature in the sonicating medium can be monitored using the temperature sensor. The
following functions can be set:
o
ff
no monitoring
stop
if the temperature falls below the limit value once again, the ultrasound operation will
be turned o
ff
. If the temperature falls below the short of the limit value once again, the
ultrasound operation will not automatically continue.
alarm when reaching or exceeding the set temperature limit value, an acoustic and visual
warning signal (red blinking LED) will be triggered. The ultrasound operation will not be
turned o
ff
.
a) Setting the type of monitoring
SET
Prog 3* Ampl
10%
Time set
00:00:15
elapsed
00:00:00
Pulsation
off
Temperature
---°C
Energy 0.000
kJ
Temperature
---°C
Prog 3* Ampl
10%
Time set
00:00:15
elapsed
00:00:00
Pulsation
off
Alarm at
0°C
---°C
Energy 0.000
kJ
Alarm at
0°C
---°C
Prog 3* Ampl
10%
Time set
00:00:15
elapsed
00:00:00
Pulsation
off
Stop at
0°C
---°C
Energy 0.000
kJ
Stop at
0°C
---°C
Example:
SET
Prog 3* Ampl
10%
Time set
00:00:15
elapsed
00:00:00
Pulsation
off
Temperature
---°C
Energy 0.000
kJ
Temperature
---°C
Prog 3* Ampl
10%
Time set
00:00:15
elapsed
00:00:00
Pulsation
off
Alarm at
0°C
---°C
Energy 0.000
kJ
Alarm at
0°C
---°C
Prog 3* Ampl
10%
Time set
00:00:15
elapsed
00:00:00
Pulsation
off
Alarm at
0°C
---°C
Energy 0.000
kJ
Alarm at
60°C
---°C
Prog 3* Ampl
10%
Time set
00:00:15
elapsed
00:00:00
Pulsation
off
Alarm at
0°C
---°C
Energy 0.000
kJ
Stop at
60°C
---°C
Prog 3* Ampl
10%
Time set
00:00:15
elapsed
00:00:00
Pulsation
off
Stop at
0°C
---°C
Energy 0.000
kJ
0°C
Prog 3* Ampl
10%
Time set
00:00:15
elapsed
00:00:00
Pulsation
off
Stop at
0°C
---°C
Energy 0.000
kJ
60°C
The maximum setting value is 120 °C, the minimum is -10 °C.
As shown in the example, the values for the Alarm and Stop functions are carried over when
de
fi
ning a limit temperature. An Alarm is triggered simultaneously in Stop mode.
Temperature measured at
temperature sensor
No sensor: ---°C
Prog 1* Ampl
20%
Time set
00:01:00
00:00:09
Pulsation 4.0s
4.0s
Temperature 31°C
Energy 0.000
kJ
SET
or
START
STOP
back to stand-by mode
Summary of Contents for Sonoplus HD4050