10 / 46
51376-010 GB/2020-10
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
1.4.4
Electromagnetic ambient conditions (EMC)
The device was tested to DIN EN 61326-1 for electromagnetic compatibility (EMC) and complies
with the requirements for class B devices according to EN 55011.
It is suitable for use in facilities and areas which are directly connected to a public low-voltage
supply network, e.g. medical laboratory facilities.
1.4.5
Remote operation
Several options are available for remote control and remote operation. Depending on
requirements, the most convenient solution may be selected.
Operating
element
Methods
Functions
Connection
1
Key
manual
Ultrasound operation ON/OFF
Fixed on the ultrasonic
converter
2
Foot switch
foot-operated Ultrasound operation ON/OFF
Remote control
connector
3
External
control signal • Ultrasound operation ON/OFF
• Status con
fi
rmation
Remote control
connector
4
External
data protocol • Status inquiry
• Full device control
RS 232 interface
a) Remote control connector
Contact assignment and functional description of the remote control connector
Diagram
Contact
Assignment
Signal
Function
1
2
3
4
5
1
Input
0V L
5V H
L
ultrasound operation ON
H
ultrasound operation OFF
2
Output
0V L
5V H
L
ultrasound operation OFF
H
ultrasound operation ON
3
Earth
4
Output
like contact 2
5
not assigned
Notes:
• The connection of power or voltage sources of any type is not permitted.
• The source resistance for the outputs is of 3.3 k
Ω
. The input resistance for subsequent signal
inputs, e.g. optocouplers, must be greater than or equal to 47 k
Ω
.
• The function of the signal input (contact 1) depends on the mode of operation that is set for
the pulsation (see chapter 3.2.2). With the operating mode "by hand key", control is conducted
statically, i.e. the ultrasound operation is always activated under L status, and deactivated under
H status.
With all other modes of operation, the input works as an edge-controlled changeover switch.
The input resistance measures approx. 50 k
Ω
.