Ultrasonic converter
Standard/booster horn
Vigorously turn sickle spanner
HS 40/42 long, in the opposite
direction
Plastic ruler
Threaded bolt
31 / 46
51376-010 GB/2020-10
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
Assembly
Keep the mating surfaces/threads to be screwed clean, using alcohol and a
fi
bre-free cloth if
needed.
Screw in approx. 2/3 of the length of the threaded bolts of the new standard/booster horn to be
mounted, into the mounting surface of the ultrasonic converter.
Place a plastic ruler or similar on the threaded portion of the bolt, and press lightly in the
direction of the arrow
this will prevent the bolt from turning any further when the standard/
booster horn is screwed on.
Screw the standard/booster horn onto the threaded bolt and
fi
rst tighten by hand after
removing the ruler.
Then, mount the standard/booster horn
fi
rmly onto the ultrasonic converter using both sickle
spanners.
Detailed information on the individual standard/booster horns can be found in the separate product
information.
Caution!
The black housing of the ultrasonic converter and the aluminium cylinder (ultrasonic oscillating
system) may not be turned against each other. The ultrasonic oscillating system and its electrical
connections would be damaged as a result.