14 / 46
51376-010 GB/2020-10
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
2.1
Scope of delivery
The scope of delivery will depend on the size of the order. However, it will generally include the
following parts:
1 Ultrasonic generator GM .... - see delivery note
1
Ultrasonic converter UW ...
1
Standard/booster horn SH ...
1
Probe
1 Tool
set
1 Instructions for Use manual
Additional accessories according to order – see delivery note
2.2
Set-up / assembly
Place the ultrasonic generator atop a
fi
rm, level and dry surface. In doing so,
•
do not group it or stack it over other electrical or electronic devices,
•
do not cover the ventilation holes on the ultrasonic converter's housing,
•
guard against moisture and wetness - risk of electric shock.
The ultrasonic generator must be positioned in such a way that disconnection from the power
supply is easily possible.
During delivery, the standard/booster horn (if required) is already
fi
rmly screwed to the
corresponding ultrasonic converter.
Other standard/booster horns are mountable. To do so, see instructions in chapter 4.4.
Screw the delivered probe to the standard/booster horn, see chapter 4.3.1.
Position the ultrasonic converter safely and correctly inside a suitable holder.
To do so, a
ffi
x the ultrasonic converter on the black housing alone.
Position the ultrasonic converter with holder in a sound proof box.
The
ultrasonic converter is connected to the ultrasonic generator through a push-pull circular
connector with lock (MINISNAP). To connect, hold the connector from the handle and position
it in such a way that the arrow points to the left. The connector can only be plugged in when in
this position. Plug the connector into the socket up to the limit, until it snaps in place. In order
to unplug the connection, hold the connector only from the handle, pull it back and remove the
connector. If necessary, see chapter 1.
Verify that the mains switch is in the "0" position.
Before the mains connection, identify the mains voltage at hand and compare it with the
voltage speci
fi
ed on the type plate on the generator bottom. Connect to a grounded socket
only if type is suitable.
Summary of Contents for Sonoplus HD4050