www.balluff.com
25
english
10
Reset all values with the calibration device (reset)
NOTICE
Interference in function
Resetting all values while the system is running may
result in malfunctions.
►
Stop the system before performing the reset.
The reset function can be used to
restore all the settings to the factory settings. For a reset
the magnet may also be located
outside the measuring range.
Activate reset
►
Hold down
and
simultaneously for at least 4 s.
> 4 s
1
2
►
Release buttons.
⇒
Reset is activated.
Reset
►
Hold down
and
for at least 4 s.
> 4 s
1
2
►
Release buttons.
⇒
All values are reset.
⇒
Current position value is displayed.
⇒
Reset is deactivated.
Abort reset
Resetting can be aborted without any changes
being saved after the Activate reset step.
►
Briefly press
and
simultaneously (< 1 s).
< 1 s
1
2
►
Release buttons.
⇒
Current position value is displayed.
1
= blue button
2
= gray button
BTL7-A/E501-M_ _ _ _-J-DEXC-TA12
Magnetostrictive Linear Position Sensor – Rod Style
Type of protection “db” and “ta”
Flameproof enclosure
Summary of Contents for BTL7-A/E501-M Series
Page 2: ...www balluff com...
Page 4: ...www balluff com...
Page 34: ......
Page 36: ...www balluff com...
Page 66: ......
Page 68: ...www balluff com...
Page 98: ......
Page 100: ...www balluff com...
Page 130: ......
Page 132: ...www balluff com...
Page 162: ......
Page 164: ...www balluff com...
Page 178: ...16 5 1 1 2 3 BTL 5 2 BTL BTL 5 BTL7 A E501 M_ _ _ _ J DEXC TA12 db ta...
Page 182: ...20 7 7 3 2 22 1 1 2 1 2 24 7 3 7 4 7 3 3 25 BTL BTL7 A E501 M_ _ _ _ J DEXC TA12 db ta...