INJECTOR CALIBRATOR CD3
MAINTENANCE AND TROUBLESHOOTING
4.2.6 Checking Main PC Board
Between these terminals
on the main P.C. board: Voltage should be:
If not, probable cause is:
________________________________________________________________________________
15 and 16
11 to 13 VAC
Defective fuse F-2 or transformer T2.
8 and 1
4.5 to 5 VDC
Defective main P.C. Board.
8 and 2
3 to 5 VDC
Defective optical detector #1.
8 and 2 (with
0 to 2 VDC
Defective optical detector #1.
detector blocked)
8 and 3 (with cam
3 to 5 VDC
Defective optical detector #2.
top dead center)
8 and 3 (with cam
0 to 2 VDC
Defective optical detector #2.
bottom dead center)
8 and 5
0 VDC
8 and 6
0 or 4.5 VDC*
Defective transducer.
8 and 7
0 or 4.5 VDC*
Defective transducer.
8 and 9
5.5 or 6.0 VDC
Defective main P.C. Board.
8 and 14 (with cam
4.5 to 5 VDC
Defective main P.C. Board.
top dead center)
8 and 14 (with cam
0 VDC
Defective main P.C. Board.
bottom dead center)
8 and 5
4.5 to 5 VDC
Defective K1 relay.
________________________________________________________________________________
*Movement of the flow transducer encoder arm will cause voltage change.
ADDITIONAL VOLTAGE CHECKS
Terminals:
Voltage should be:
If not, probable cause is:
________________________________________________________________________________
across K1 coil
115 VAC
Defective motor start switch or S6
M1 auxiliary contact.
L6 or L7 solenoid on
should be energized
Defective K3 (or when K3 is
manually transducer operated,
they should not be energized).
_________________________________________________________________________________
Summary of Contents for 67-7622
Page 26: ...Page 2 8 Bacharach Inc Instructions 67 9319 INSTALLATION INJECTOR CALIBRATOR CD3 NOTES ...
Page 54: ...Page 3 28 Bacharach Inc Instructions 67 9319 OPERATION INJECTOR CALIBRATOR CD3 NOTES ...
Page 96: ...Page A 2 Bacharach Inc Instructions 67 9319 APPENDIX INJECTOR CALIBRATOR CD3 NOTES ...