Page 3-22
Bacharach, Inc.
Instructions 67-9319
OPERATION
INJECTOR CALIBRATOR CD3
Figure 3-47. Injector Setup
Figure 3-48. Injector Clamping
Figure 3-49. Timing Light & Clamp Monitor
The air cylinders will retract and clamp injector.
11. Press the BODY TO SEAT CLAMP button.
12. Release BODY TO SEAT CLAMP button when
the BODY TO SEAT monitor turns green.
3.2.5.4 Fuel Inlet
13. Disconnect the fuel inlet tubing from the fuel
arm by unscrewing the knurled knob.
14. Swing the fuel arm back toward console.
15. Connect fuel inlet fitting to manifold INLET.
3.2.5.5 Injector
16. Lay the injector into the support.
17. Guide the spring and plunger of the injector
into the manifold block.
3.2.5.6 Injector Clamping
18. Push and hold the HOLD TO OPERATE button
with left hand.
19. While holding down the HOLD TO OPERATE
button, move the CLAMP/UNCLAMP lever to
the CLAMP position with right hand.
Maximum clamping force shown on the CLAMP
MONITOR/TIMING PLUNGER TO BODY gauge
will be in the green segment of the dial and the
PLUNGER TO BODY timing light will be lit.
20. Wait until the PLUNGER TO BODY timing
light goes out (after about 3 seconds).
21. Release the HOLD TO OPERATE button
first
and
then
the CLAMP/UNCLAMP lever. The
lever will return to the center position.
For clamping to occur correctly, you must follow
this sequence carefully.
The gauge reading will decrease from the green
segment of the dial to a lower reading (red seg-
ment) immediately after the CLAMP/UNCLAMP
valve lever is returned to the center position.
The injector is now ready for testing (Section 3.2.7).
15
Manifold Inlet
5 Discharge Head
16
Injector
6 Flowmeter Tubing
19
CLAMP
18
HOLD TO OPERATE
19
Plunger to Body
Clamp Monitor
20
Plunger to Body
Timing Light