0019-9376
Revision 0
14
Monoxor
®
XR User Manual
Specification
Description
Temperature
Storage: -20° to 50° C
( -4° to 122° F )
| 0° to 20° C
( 32° to 68° F )
optimal
Operation: -5° to 45° C (23° to 113° F)
Reference: 20° ± 2° C (68° ± 4° F)
Humidity
Storage: 15 to 90% RH, non-condensing
Operation: 15 to 95% RH, non-condensing
Reference: 45 ± 10% RH, non-condensing
Pressure
1 atmosphere ± 10%
Weight
16 ounces
(454 g)
with batteries
Dimensions
(H × W × D)
8.0” × 3.6” × 2.3”
(20.3 cm × 9.1 cm × 5.8 cm)
Warm-up Time
60 seconds
Gas Sample Flow Rate
300 to 700 cc/min
Sensors
CO Sensor: Electrochemical
(P/N: 0024-0997)
B-Smart CO Sensor: Electrochemical
(P/N: 0024-1795)
Temperature: K-Type thermocouple
Product Approvals and
Regulatory Compliance
EN50270: 2015
(CE Mark)
EMC tested in accordance with European Directive
RoHS Compliance
Case Construction
High impact ABS plastic with co-molded rubber.
Optional protective rubber boot with molded-in magnets.
Display
Monochrome with backlight
USB Connector
Mini B
(USB 2.0)
Memory
100 locations for storing test results
IrDA Port
Protocol: IrDA-SIR Data Bits: 8 Stop Bits: 1
Baud Rate: 9600 Parity: None
Power Supply Options
Batteries (4 AA)
Type:
Disposable Alkaline
(Included)
Duration:
15 hours min, continuous max draw
Type:
Disposable Lithium
Duration:
20 hours, continuous max draw
Type:
Rechargeable
Duration:
8 hours, continuous max draw