0019-9376
Revision 0
30
Monoxor
®
XR User Manual
Setup Menu
Function
The Language Selection option allows the user to choose a language
for all menus. Use the UP ( ) and DOWN ( ) arrow buttons to scroll
through language options. Use ENTER to enable the selected language.
Three languages are available: English, French, and Spanish.
The button sound option is used to select whether or not the
analyzer makes an audible sound every time a key is pressed. Select
OFF to disable this feedback, or ON to enable this feedback. Then
press ENTER to select, or ESC to discard changes.
The CAL Reminder Period option sets a time period after which the
analyzer displays a calibration reminder message during warm-up.
Calibration reminders can be set to occur never, 6, 8, 10, 12, or 15
months after the last calibration. Select NEVER
(the default setting)
to disable this feature. When the preset period is exceeded the
instrument will display the reminder, and how long since the sensor
was last calibrated. If a calibration reminder is displayed, the operator
can press the RUN/HOLD key to move to the RUN/HOLD Screen for
normal operation. Regular calibration periods of 6 months to 1 year
are recommended.
Set the calibration reminder period as follows:
1. Use the UP ( ) and DOWN ( ) arrow buttons to select the
desired time period.
2. Press ENTER to save the selection or ESC to revert to the
previous setting.