Monoxor
®
XR User Manual
3.4 Ambient CO Menu
Ambient CO
Function
Access the Ambient CO Menu.
When initiated, the Ambient CO feature monitors CO values
continuously and captures a reading every minute for 15 minutes
(a
total of 16 readings from t
0
to t
15
).
Press ENTER to initiate the Ambient CO test. This begins a 15-minute
test cycle, during which a status screen is displayed. It shows the
starting ambient CO value, the current CO value, and the elapsed
time into the test.
NOTE:
Press the F2 key to cancel a test
in progress.
After the test is complete, the Ambient CO Summary screen
is displayed. This is a scrollable window that shows the 16 CO
“snapshot” readings, as well as the maximum CO reading that was
sampled during the entire test.
NOTE:
The Max CO Reading is the highest
sampled CO reading – even if the reading
was taken in between one of the sample
“snapshot” readings.