0019-9376
Revision 0
35
Monoxor
®
XR User Manual
3.8 Calibration Menu
Calibration Menu
Function
Calibration is performed by applying known values and accessing
the password-protected menu items. When the Calibration Menu is
selected, the user must enter a 4-digit numeric security code in order
to proceed to the calibration options. The default password is 1111.
Use the UP ( ) and DOWN ( ) arrow buttons to scroll through
numerals 0-9 until the desired numeral is reached. Press ENTER to
advance to the next position of the password. Press ENTER after all
four digits are set. Press ESC to return to the SETUP MENU.
NOTE:
The calibration password can be changed
through the Fyrite
®
User Software (FUS).
Refer to Chapter 5 for additional screens and calibration procedures.