0019-9376
Revision 0
56
Monoxor
®
XR User Manual
4. Set thermocouple simulator to 32° F (0° C), and then use the UP ( ) and DOWN ( ), LEFT
( ) and RIGHT ( ) arrow buttons to enter an Applied value that exactly equals the setting
of the simulator.
Alternatively:
Submerge probe tip into an ice-water bath with a thermometer, wait several
minutes, and then use the UP ( ) and DOWN ( ) arrow buttons to enter an Applied value that
exactly equals the thermometer reading.
NOTE:
The calibration range is from 32 to 41° F
(0 to 5° C)
. An attempt to
calibrate outside this range will cause the message “Applied Value High”
(or
Low)
to appear at the bottom of the screen.
5. Wait until the Measured reading stabilizes, and then press ENTER to calibrate the TS-Zero
Measured value to that of the Applied value, after which the message “Good Calibration”
should briefly appear followed by the CALIBRATE TS-SPAN screen.
6. Set thermocouple simulator to 572° F
(300° C)
, and then use the UP ( ) and DOWN ( ),
LEFT ( ) and RIGHT ( ) arrow buttons to enter an Applied value that exactly equals the
setting of the simulator.
Figure 5-6:
Calibration T-Span Screen
Alternatively:
Submerge probe tip into a container of boiling water with a thermometer, wait
several minutes, and then use the arrow buttons to enter an Applied value that exactly equals
the thermometer reading.
NOTE:
The calibration range is from 175 to 625° F
(80 to 329° C)
. An attempt
to calibrate outside this range will cause the message “Applied Value High”
(or
Low)
to appear at the bottom of the screen.
7. Wait until the Measured reading stabilizes, and then press ENTER to calibrate the TS-Span
Measured value to that of the “Applied” value, after which the message “Good Calibration”
should briefly appear followed by the CALIBRATION menu being re-displayed.