4000
4-/5-Gas Measuring Instruments Unit
Service Manual
5-23
Primary error symptom Primary cause
(of error)
Secondary cause of
error
Possible tests/
establishing cause
Remedy
Plug and/or cable
faulty or
disconnected
Not connected or
irreparably
damaged when O
2
sensor replaced
Check:
service mode and
on O
2
sensor,
DSM=07 after
zeroing?
O2=0: no signal
Connect or replace
cable
Cable to measuring
instruments unit
disconnected
Measuring
instruments unit
cable disconnected
when O
2
sensor
replaced
When cable
connected and OK:
approx. 40 k
Ω
internal resistance
should be
measurable
Connect or replace
cable
O
2
sensor faulty
Condensate in
sensor due to
operation at below
5° C
Storage or operation
at temperatures
above 50° C
Stored at below 0° C
for a long period
(i.e. days)
Sensor voltage
lower than 5 mV
Ambient tempera-
ture fluctuating too
much
Condensate in hose?
Ambient/operating
conditions?
Sensor is too old:
sensor voltage must
be between 5 and
14 mV (see raw
values)
Replace sensor,
ensure required
operating conditions
Leak in exhaust
pipe
(Applies to 4-stroke
engines only and
lambda setpoint
value
≤
1)
CO + CO
2
measurement value
less than 15 % by
volume?
O
2
> 0 %
Repair leak in
exhaust
Leak in instrument (Applies to 4-stroke
engines only and
lambda setpoint
value
≤
1)
CO + CO
2
measurement value
less than 15 % by
volume?
O
2
> 0 %, carry out
leak test
Find leak and repair
Hose disconnected
after O2 sensor
Possible leak
between measuring
instruments unit
and O
2
sensor
At least 10 cm hose
is necessary after O
2
sensor (465)
Connect hose to
outlet or repair leak
Software
interference due to
EMC
Saved starting value
has been deleted
Service mode: when
setpoint 7 to 14 mV
in air, sensor OK
Initialise sensor again
in service mode
Oxygen measurement
value incorrect and O
2
sensor faulty
Measuring
instruments unit
faulty
Service mode: ES,
SZ not zero?
Replace measuring
instruments unit
Summary of Contents for AVL DiCom 4000
Page 4: ......
Page 6: ......
Page 22: ...Periodic Maintenance AVL 4000 1 12 Service Manual ...
Page 47: ...4000 Evaluation Unit Service Manual 4 15 4 9 Main Board 4 9 1 Components Location Diagram ...
Page 48: ...Evaluation Unit AVL 4000 4 16 Service Manual 4 9 2 Circuit Diagrams ...
Page 49: ...4000 Evaluation Unit Service Manual 4 17 ...
Page 50: ...Evaluation Unit AVL 4000 4 18 Service Manual ...
Page 51: ...4000 Evaluation Unit Service Manual 4 19 ...
Page 52: ...Evaluation Unit AVL 4000 4 20 Service Manual ...
Page 54: ...Evaluation Unit AVL 4000 4 22 Service Manual 4 10 2 Circuit Diagrams ...
Page 55: ...4000 Evaluation Unit Service Manual 4 23 ...
Page 56: ...Evaluation Unit AVL 4000 4 24 Service Manual ...
Page 57: ...4000 Evaluation Unit Service Manual 4 25 ...
Page 59: ...4000 Evaluation Unit Service Manual 4 27 4 11 2 Circuit Diagram ...
Page 61: ...4000 Evaluation Unit Service Manual 4 29 4 12 2 Circuit Diagram ...
Page 63: ...4000 Evaluation Unit Service Manual 4 31 4 13 2 Circuit Diagram ...
Page 65: ...4000 Evaluation Unit Service Manual 4 33 4 14 2 Circuit Diagram ...
Page 67: ...4000 Evaluation Unit Service Manual 4 35 4 15 2 Circuit Diagram ...
Page 69: ...4000 Evaluation Unit Service Manual 4 37 4 16 2 Circuit Diagram ...
Page 71: ...4000 Evaluation Unit Service Manual 4 39 4 17 2 Circuit Diagram ...
Page 73: ...4000 Evaluation Unit Service Manual 4 41 4 18 2 Circuit Diagram ...
Page 75: ...4000 Evaluation Unit Service Manual 4 43 4 19 2 Circuit Diagram ...
Page 77: ...4000 Evaluation Unit Service Manual 4 45 4 20 2 Circuit Diagram ...
Page 79: ...4000 Evaluation Unit Service Manual 4 47 4 21 2 Circuit Diagram ...
Page 124: ...4 5 Gas Measuring Instruments Unit AVL 4000 5 28 Service Manual ...
Page 149: ...4000 Opacity Measuring Chamber Type A Service Manual 6 25 Fig 6 30 Fig 6 31 ...
Page 150: ...Opacity Measuring Chamber Type A AVL 4000 6 26 Service Manual Fig 6 32 ...
Page 180: ...Spare Parts Lists AVL 4000 8 14 Service Manual ...
Page 182: ...Tools for the Service Technician AVL 4000 9 2 Service Manual ...