4000
Evaluation Unit
Service Manual
4-57
Field 4
Measuring instrument unit status and errors
SS
System status
Message
08
Processor in warmup phase
other messages are irrelevant
ZS
Zero status
This tells you which warmup phase the tester is in.
Message
0
wait for obligatory zero calibration
1
1 minute mode
5
5 minute mode
E
15 minute mode
1E
30 minute mode
OS
Output status
irrelevant
ES
Error status
When the tester is switched on, the last two statuses usually display messages 30, 20 or
10. These indicate the warm-up phase. After approx.7min, the display should read 00.
Field 5
(Depending on the software version not all of the following values may be displayed.)
1
st
line
date
2
nd
line
time
3
rd
and 4
th
line software version
5
th
and 6
th
line silicon serial no. of mainboard
7
th
line
serial no. of measuring instrument unit
(not including pneumatic components module)
Field 6
Software version installed in the measuring instrument unit
Version "CC 31" (or "D841" with the NO option) must always be displayed.
Field 7
Pneumatic components status field
Pump
pump status (0 = off, 1 = on)
Operating hours counter
digital pressure value during the leak test
SOL 1 and 2:
valve settings (0 = valve off, 1 = valve on)
on
off
valve 1
ZERO
Sample
valve 2
Backflush valve
Sample
below:
voltage input signal
Summary of Contents for AVL DiCom 4000
Page 4: ......
Page 6: ......
Page 22: ...Periodic Maintenance AVL 4000 1 12 Service Manual ...
Page 47: ...4000 Evaluation Unit Service Manual 4 15 4 9 Main Board 4 9 1 Components Location Diagram ...
Page 48: ...Evaluation Unit AVL 4000 4 16 Service Manual 4 9 2 Circuit Diagrams ...
Page 49: ...4000 Evaluation Unit Service Manual 4 17 ...
Page 50: ...Evaluation Unit AVL 4000 4 18 Service Manual ...
Page 51: ...4000 Evaluation Unit Service Manual 4 19 ...
Page 52: ...Evaluation Unit AVL 4000 4 20 Service Manual ...
Page 54: ...Evaluation Unit AVL 4000 4 22 Service Manual 4 10 2 Circuit Diagrams ...
Page 55: ...4000 Evaluation Unit Service Manual 4 23 ...
Page 56: ...Evaluation Unit AVL 4000 4 24 Service Manual ...
Page 57: ...4000 Evaluation Unit Service Manual 4 25 ...
Page 59: ...4000 Evaluation Unit Service Manual 4 27 4 11 2 Circuit Diagram ...
Page 61: ...4000 Evaluation Unit Service Manual 4 29 4 12 2 Circuit Diagram ...
Page 63: ...4000 Evaluation Unit Service Manual 4 31 4 13 2 Circuit Diagram ...
Page 65: ...4000 Evaluation Unit Service Manual 4 33 4 14 2 Circuit Diagram ...
Page 67: ...4000 Evaluation Unit Service Manual 4 35 4 15 2 Circuit Diagram ...
Page 69: ...4000 Evaluation Unit Service Manual 4 37 4 16 2 Circuit Diagram ...
Page 71: ...4000 Evaluation Unit Service Manual 4 39 4 17 2 Circuit Diagram ...
Page 73: ...4000 Evaluation Unit Service Manual 4 41 4 18 2 Circuit Diagram ...
Page 75: ...4000 Evaluation Unit Service Manual 4 43 4 19 2 Circuit Diagram ...
Page 77: ...4000 Evaluation Unit Service Manual 4 45 4 20 2 Circuit Diagram ...
Page 79: ...4000 Evaluation Unit Service Manual 4 47 4 21 2 Circuit Diagram ...
Page 124: ...4 5 Gas Measuring Instruments Unit AVL 4000 5 28 Service Manual ...
Page 149: ...4000 Opacity Measuring Chamber Type A Service Manual 6 25 Fig 6 30 Fig 6 31 ...
Page 150: ...Opacity Measuring Chamber Type A AVL 4000 6 26 Service Manual Fig 6 32 ...
Page 180: ...Spare Parts Lists AVL 4000 8 14 Service Manual ...
Page 182: ...Tools for the Service Technician AVL 4000 9 2 Service Manual ...