Primary error symptom Primary cause
(of error)
Secondary cause of
error
Possible tests/
establishing cause
Remedy
Warm-up error
FE, SE, SZ stable at
128 ± 10?
ES=000000 ?
Wait until instrument
has warmed up (SS
must be 00)
Replace measuring
instruments unit
Detector preamplifier faulty or feedback
control inaccurate
FE, SE, SZ stable at
128 ± 10?
ES=000000?
Reset, or measuring
instruments unit
faulty
Replace detector
Replace measuring
instruments unit
Hoses inside are disconnected or faulty Test flow rates
Replace hoses
Measurement values
fluctuating
significantly after
zeroing or calibration
Exhaust gas, petrol or other gases in the
air
Can you smell
anything?
Air the room or
connect hose on
carbon filter to fresh
air supply
Substantial
fluctuations in
measurement
values
See “Measurement
values fluctuating
significantly”
Ambient air heavily polluted by petrol,
exhaust gas…
Can you small
anything?
Negative
measurement
values?
Air the room or
connect hose on
carbon filter to fresh
air supply
ZEROING/ZERO not
possible
EMC interference
during zeroing
Check error status
Switch instrument
OFF/ON
Carry out power
reset; then
CALIBRATION
Summary of Contents for AVL DiCom 4000
Page 4: ......
Page 6: ......
Page 22: ...Periodic Maintenance AVL 4000 1 12 Service Manual ...
Page 47: ...4000 Evaluation Unit Service Manual 4 15 4 9 Main Board 4 9 1 Components Location Diagram ...
Page 48: ...Evaluation Unit AVL 4000 4 16 Service Manual 4 9 2 Circuit Diagrams ...
Page 49: ...4000 Evaluation Unit Service Manual 4 17 ...
Page 50: ...Evaluation Unit AVL 4000 4 18 Service Manual ...
Page 51: ...4000 Evaluation Unit Service Manual 4 19 ...
Page 52: ...Evaluation Unit AVL 4000 4 20 Service Manual ...
Page 54: ...Evaluation Unit AVL 4000 4 22 Service Manual 4 10 2 Circuit Diagrams ...
Page 55: ...4000 Evaluation Unit Service Manual 4 23 ...
Page 56: ...Evaluation Unit AVL 4000 4 24 Service Manual ...
Page 57: ...4000 Evaluation Unit Service Manual 4 25 ...
Page 59: ...4000 Evaluation Unit Service Manual 4 27 4 11 2 Circuit Diagram ...
Page 61: ...4000 Evaluation Unit Service Manual 4 29 4 12 2 Circuit Diagram ...
Page 63: ...4000 Evaluation Unit Service Manual 4 31 4 13 2 Circuit Diagram ...
Page 65: ...4000 Evaluation Unit Service Manual 4 33 4 14 2 Circuit Diagram ...
Page 67: ...4000 Evaluation Unit Service Manual 4 35 4 15 2 Circuit Diagram ...
Page 69: ...4000 Evaluation Unit Service Manual 4 37 4 16 2 Circuit Diagram ...
Page 71: ...4000 Evaluation Unit Service Manual 4 39 4 17 2 Circuit Diagram ...
Page 73: ...4000 Evaluation Unit Service Manual 4 41 4 18 2 Circuit Diagram ...
Page 75: ...4000 Evaluation Unit Service Manual 4 43 4 19 2 Circuit Diagram ...
Page 77: ...4000 Evaluation Unit Service Manual 4 45 4 20 2 Circuit Diagram ...
Page 79: ...4000 Evaluation Unit Service Manual 4 47 4 21 2 Circuit Diagram ...
Page 124: ...4 5 Gas Measuring Instruments Unit AVL 4000 5 28 Service Manual ...
Page 149: ...4000 Opacity Measuring Chamber Type A Service Manual 6 25 Fig 6 30 Fig 6 31 ...
Page 150: ...Opacity Measuring Chamber Type A AVL 4000 6 26 Service Manual Fig 6 32 ...
Page 180: ...Spare Parts Lists AVL 4000 8 14 Service Manual ...
Page 182: ...Tools for the Service Technician AVL 4000 9 2 Service Manual ...