4000
Evaluation Unit
Service Manual
4-11
4.2.1.10 Memory Adapter
The Memory Adapter (AVL identification no. BB0730) is installed on the main board.
•
Insert the Memory Adapter at slots J8 and J9 on the main board
•
Mark the memory adapter option on the options sticker.
4.2.1.11 Chamber Adapter (AVL DiCom 4000)
The Chamber Adapter is needed to operate the opacity measuring chamber.
4.2.1.12 Extension Slot
There is a free extension card slot for future upgrading as the result of product
development. An option installed in this slot is located above the Speed/Angle Board.
If an extension card is installed, the side panel of the housing has to be changed.
4.2.1.13 Mains Switch Assembly
The mains switch assembly consists of the mains cable connection as an inlet connector
for non-heating appliances, the power switch, externally accessed mains fuse and a filter
required by the statutory provisions for mains pollution.
4.2.1.14 Converting 230 V to 115 V Device
•
Replace 230 V pump (on the real panel) with 115 V pump
The hose lengths must not be changed!
•
Both fuses in the mains switch group must be replaced by 3.15 A slow-blow fuses.
•
The modification from 230 V to 115 V must be indicated clearly on the rating plate
(with a warning on a sticker)
The power supply unit can continue to be used without any modifications.
4.2.1.15 Power Supply Unit
All members of the instrument family contain the same purchasable power supply unit.
Technical data
−
Computer Products NFS80-7606
−
long-range input 90…260V AC, 125…360V DC
−
outputs (short-circuit-proof):
+5V / 8A (±2%, ripple 1%)
for digital part
+15V / 2.5A (±3%, ripple 1%)
mainly analogue supply
-15V /2.5A (±3%, ripple 1%)
mainly analogue supply
+24V / 2A (+10/-5%, ripple 1%) power (heating, fan, solenoid valve,
strobe,…)
−
Output: 80 W (110 W with forced cooling)
−
UL, CSA, VDE, IEC1010, IEC 950 approved
−
mains pollution under VDE limit B
Summary of Contents for AVL DiCom 4000
Page 4: ......
Page 6: ......
Page 22: ...Periodic Maintenance AVL 4000 1 12 Service Manual ...
Page 47: ...4000 Evaluation Unit Service Manual 4 15 4 9 Main Board 4 9 1 Components Location Diagram ...
Page 48: ...Evaluation Unit AVL 4000 4 16 Service Manual 4 9 2 Circuit Diagrams ...
Page 49: ...4000 Evaluation Unit Service Manual 4 17 ...
Page 50: ...Evaluation Unit AVL 4000 4 18 Service Manual ...
Page 51: ...4000 Evaluation Unit Service Manual 4 19 ...
Page 52: ...Evaluation Unit AVL 4000 4 20 Service Manual ...
Page 54: ...Evaluation Unit AVL 4000 4 22 Service Manual 4 10 2 Circuit Diagrams ...
Page 55: ...4000 Evaluation Unit Service Manual 4 23 ...
Page 56: ...Evaluation Unit AVL 4000 4 24 Service Manual ...
Page 57: ...4000 Evaluation Unit Service Manual 4 25 ...
Page 59: ...4000 Evaluation Unit Service Manual 4 27 4 11 2 Circuit Diagram ...
Page 61: ...4000 Evaluation Unit Service Manual 4 29 4 12 2 Circuit Diagram ...
Page 63: ...4000 Evaluation Unit Service Manual 4 31 4 13 2 Circuit Diagram ...
Page 65: ...4000 Evaluation Unit Service Manual 4 33 4 14 2 Circuit Diagram ...
Page 67: ...4000 Evaluation Unit Service Manual 4 35 4 15 2 Circuit Diagram ...
Page 69: ...4000 Evaluation Unit Service Manual 4 37 4 16 2 Circuit Diagram ...
Page 71: ...4000 Evaluation Unit Service Manual 4 39 4 17 2 Circuit Diagram ...
Page 73: ...4000 Evaluation Unit Service Manual 4 41 4 18 2 Circuit Diagram ...
Page 75: ...4000 Evaluation Unit Service Manual 4 43 4 19 2 Circuit Diagram ...
Page 77: ...4000 Evaluation Unit Service Manual 4 45 4 20 2 Circuit Diagram ...
Page 79: ...4000 Evaluation Unit Service Manual 4 47 4 21 2 Circuit Diagram ...
Page 124: ...4 5 Gas Measuring Instruments Unit AVL 4000 5 28 Service Manual ...
Page 149: ...4000 Opacity Measuring Chamber Type A Service Manual 6 25 Fig 6 30 Fig 6 31 ...
Page 150: ...Opacity Measuring Chamber Type A AVL 4000 6 26 Service Manual Fig 6 32 ...
Page 180: ...Spare Parts Lists AVL 4000 8 14 Service Manual ...
Page 182: ...Tools for the Service Technician AVL 4000 9 2 Service Manual ...