4000
4-/5-Gas Measuring Instruments Unit
Service Manual
5-17
Primary error symptom Primary cause
(of error)
Secondary cause of
error
Possible tests/
establishing cause
Remedy
Measurement cell
contaminated with
condensate
Condensate end of
pump faulty
Condensate
discharge hose
clamped closed
Water, condensate
escaping at oxygen
sensor
Clean or replace
measurement cell
Replace measuring
instruments unit
Error in measuring
instruments unit
Hardware fault or
software error
Incorrect control
parameters
Faulty EPROMs or
processor
Check error status
ES and zero status
ZS, ES=000000 = ok
ZS=01, 05, 0F or 1E
after zeroing = ok
Try reset
Try reset or replace
measuring
instruments unit
Pump cable faulty
(465)
Interference, self-test Permanently zero
Use edge protectors
(465)
Ambient
temperature too
cold (less than
5° C)
Instrument stabilises
itself every 15 min
Is the instrument set
up out-of-doors?
Place instrument
somewhere else
Ambient temperature fluctuates
significantly (more than 15° C)
Instrument set up
near garage doors?
Place instrument
somewhere else
Low flow occurred Probe hose kinked;
did someone tread
on hose?
Zeroing occurs after
"Low Flow" error
Lay the hose without
kinks
Replace filter
Clean metal filter
(465)
Error occurred
during zeroing
Hose kinked/
blocked
O
2
signal incorrect,
Pressure value
incorrect
Service mode:
raw values O
2
:
210 to 240?
Pressure correct?
approx. 760 torr at
sea level
Check hoses
Check flow rate (at
least 4
min
l
total)
Try reset or replace
faulty part
Replace O
2
sensor
Check hoses
Fluctuating
measurement
values
Measuring
instruments unit
detector faulty
Service mode:
fluctuating CO,
CO
2
, HC values?
FE not 128 ± 10
Try reset or replace
measuring
instruments unit, see
also “Fluctuating
measurement values”
Fluctuating
measurement
values
Exhaust gas or
petrol in the air?
Can you smell
anything?
Air the room, replace
carbon filter if
necessary
Permanently
stabilising
Faulty capacitor
(465)
Smell of burning?
Replace board
Summary of Contents for AVL DiCom 4000
Page 4: ......
Page 6: ......
Page 22: ...Periodic Maintenance AVL 4000 1 12 Service Manual ...
Page 47: ...4000 Evaluation Unit Service Manual 4 15 4 9 Main Board 4 9 1 Components Location Diagram ...
Page 48: ...Evaluation Unit AVL 4000 4 16 Service Manual 4 9 2 Circuit Diagrams ...
Page 49: ...4000 Evaluation Unit Service Manual 4 17 ...
Page 50: ...Evaluation Unit AVL 4000 4 18 Service Manual ...
Page 51: ...4000 Evaluation Unit Service Manual 4 19 ...
Page 52: ...Evaluation Unit AVL 4000 4 20 Service Manual ...
Page 54: ...Evaluation Unit AVL 4000 4 22 Service Manual 4 10 2 Circuit Diagrams ...
Page 55: ...4000 Evaluation Unit Service Manual 4 23 ...
Page 56: ...Evaluation Unit AVL 4000 4 24 Service Manual ...
Page 57: ...4000 Evaluation Unit Service Manual 4 25 ...
Page 59: ...4000 Evaluation Unit Service Manual 4 27 4 11 2 Circuit Diagram ...
Page 61: ...4000 Evaluation Unit Service Manual 4 29 4 12 2 Circuit Diagram ...
Page 63: ...4000 Evaluation Unit Service Manual 4 31 4 13 2 Circuit Diagram ...
Page 65: ...4000 Evaluation Unit Service Manual 4 33 4 14 2 Circuit Diagram ...
Page 67: ...4000 Evaluation Unit Service Manual 4 35 4 15 2 Circuit Diagram ...
Page 69: ...4000 Evaluation Unit Service Manual 4 37 4 16 2 Circuit Diagram ...
Page 71: ...4000 Evaluation Unit Service Manual 4 39 4 17 2 Circuit Diagram ...
Page 73: ...4000 Evaluation Unit Service Manual 4 41 4 18 2 Circuit Diagram ...
Page 75: ...4000 Evaluation Unit Service Manual 4 43 4 19 2 Circuit Diagram ...
Page 77: ...4000 Evaluation Unit Service Manual 4 45 4 20 2 Circuit Diagram ...
Page 79: ...4000 Evaluation Unit Service Manual 4 47 4 21 2 Circuit Diagram ...
Page 124: ...4 5 Gas Measuring Instruments Unit AVL 4000 5 28 Service Manual ...
Page 149: ...4000 Opacity Measuring Chamber Type A Service Manual 6 25 Fig 6 30 Fig 6 31 ...
Page 150: ...Opacity Measuring Chamber Type A AVL 4000 6 26 Service Manual Fig 6 32 ...
Page 180: ...Spare Parts Lists AVL 4000 8 14 Service Manual ...
Page 182: ...Tools for the Service Technician AVL 4000 9 2 Service Manual ...