5-13
N.B.
Large amounts of deposits tend to be washed away when you first start
cleaning. Do not therefore feed the cleaning fluid back into the circuit to
start with. The length of time depends on the level of soiling; guideline
~30 s).
•
Switch the instrument on and activate the pump.
•
The cleaning process itself should take about half an hour
•
When cleaning is completed, rinse the system in the same way with cold water.
•
Remove the cleaning hose from the check valve and let the pump run on for ~30 s.
•
Switch the pump off.
•
Remove the cleaning hose and the probe hose.
•
Clean the probe hose and probe with clean, oil-free compressed air.
Caution: the probe hose must not be connected to the exhaust gas tester!
•
Replace the condensate filter on the rear panel of the instrument.
•
Connect the hose between the condensate filter and check valve.
•
Replace the filter in the probe hose.
•
After cleaning, carry out a leak check.
Summary of Contents for AVL DiCom 4000
Page 4: ......
Page 6: ......
Page 22: ...Periodic Maintenance AVL 4000 1 12 Service Manual ...
Page 47: ...4000 Evaluation Unit Service Manual 4 15 4 9 Main Board 4 9 1 Components Location Diagram ...
Page 48: ...Evaluation Unit AVL 4000 4 16 Service Manual 4 9 2 Circuit Diagrams ...
Page 49: ...4000 Evaluation Unit Service Manual 4 17 ...
Page 50: ...Evaluation Unit AVL 4000 4 18 Service Manual ...
Page 51: ...4000 Evaluation Unit Service Manual 4 19 ...
Page 52: ...Evaluation Unit AVL 4000 4 20 Service Manual ...
Page 54: ...Evaluation Unit AVL 4000 4 22 Service Manual 4 10 2 Circuit Diagrams ...
Page 55: ...4000 Evaluation Unit Service Manual 4 23 ...
Page 56: ...Evaluation Unit AVL 4000 4 24 Service Manual ...
Page 57: ...4000 Evaluation Unit Service Manual 4 25 ...
Page 59: ...4000 Evaluation Unit Service Manual 4 27 4 11 2 Circuit Diagram ...
Page 61: ...4000 Evaluation Unit Service Manual 4 29 4 12 2 Circuit Diagram ...
Page 63: ...4000 Evaluation Unit Service Manual 4 31 4 13 2 Circuit Diagram ...
Page 65: ...4000 Evaluation Unit Service Manual 4 33 4 14 2 Circuit Diagram ...
Page 67: ...4000 Evaluation Unit Service Manual 4 35 4 15 2 Circuit Diagram ...
Page 69: ...4000 Evaluation Unit Service Manual 4 37 4 16 2 Circuit Diagram ...
Page 71: ...4000 Evaluation Unit Service Manual 4 39 4 17 2 Circuit Diagram ...
Page 73: ...4000 Evaluation Unit Service Manual 4 41 4 18 2 Circuit Diagram ...
Page 75: ...4000 Evaluation Unit Service Manual 4 43 4 19 2 Circuit Diagram ...
Page 77: ...4000 Evaluation Unit Service Manual 4 45 4 20 2 Circuit Diagram ...
Page 79: ...4000 Evaluation Unit Service Manual 4 47 4 21 2 Circuit Diagram ...
Page 124: ...4 5 Gas Measuring Instruments Unit AVL 4000 5 28 Service Manual ...
Page 149: ...4000 Opacity Measuring Chamber Type A Service Manual 6 25 Fig 6 30 Fig 6 31 ...
Page 150: ...Opacity Measuring Chamber Type A AVL 4000 6 26 Service Manual Fig 6 32 ...
Page 180: ...Spare Parts Lists AVL 4000 8 14 Service Manual ...
Page 182: ...Tools for the Service Technician AVL 4000 9 2 Service Manual ...