4000
Error Messages and Troubleshooting
Service Manual
2-5
2.2
Opacity Measuring Chamber
2.2.1
Measuring Chamber Too Hot/Too Cold
If the error message "Measuring chamber too hot" occurs during operation, remove the
probe from the exhaust pipe and wait until the temperature returns to the permitted
range. Acknowledge the message by pressing
>>
.
If the error message "Measuring chamber too cold" is displayed, the chamber has been
cooled by cold exhaust gas. Wait until the measuring chamber has heated up again to at
least 80° C and acknowledge the message by pressing
>>
.
If the measuring chamber temperature does not return to the permitted range after
several attempts (2 to 3 times), there must be a fault.
•
If the temperature is still too low, one of the heating blocks has probably failed (see
Operating Manual - Heating Function Test or section 6.6 Heating Blocks and Orifice
Plates) or a temperature sensor has failed (see section 6.8 Temperature Sensor).
•
If the temperature is still too high, the feedback control system is malfunctioning (see
section. 4.2.1 Main Board). But check the temperature sensor as well, just in case.
2.2.2
Changeover Valve Error
This message suggests failure of the changeover valve in the opacity measuring chamber.
•
Replace changeover valve.
Do not lubricate the changeover valve.
Summary of Contents for AVL DiCom 4000
Page 4: ......
Page 6: ......
Page 22: ...Periodic Maintenance AVL 4000 1 12 Service Manual ...
Page 47: ...4000 Evaluation Unit Service Manual 4 15 4 9 Main Board 4 9 1 Components Location Diagram ...
Page 48: ...Evaluation Unit AVL 4000 4 16 Service Manual 4 9 2 Circuit Diagrams ...
Page 49: ...4000 Evaluation Unit Service Manual 4 17 ...
Page 50: ...Evaluation Unit AVL 4000 4 18 Service Manual ...
Page 51: ...4000 Evaluation Unit Service Manual 4 19 ...
Page 52: ...Evaluation Unit AVL 4000 4 20 Service Manual ...
Page 54: ...Evaluation Unit AVL 4000 4 22 Service Manual 4 10 2 Circuit Diagrams ...
Page 55: ...4000 Evaluation Unit Service Manual 4 23 ...
Page 56: ...Evaluation Unit AVL 4000 4 24 Service Manual ...
Page 57: ...4000 Evaluation Unit Service Manual 4 25 ...
Page 59: ...4000 Evaluation Unit Service Manual 4 27 4 11 2 Circuit Diagram ...
Page 61: ...4000 Evaluation Unit Service Manual 4 29 4 12 2 Circuit Diagram ...
Page 63: ...4000 Evaluation Unit Service Manual 4 31 4 13 2 Circuit Diagram ...
Page 65: ...4000 Evaluation Unit Service Manual 4 33 4 14 2 Circuit Diagram ...
Page 67: ...4000 Evaluation Unit Service Manual 4 35 4 15 2 Circuit Diagram ...
Page 69: ...4000 Evaluation Unit Service Manual 4 37 4 16 2 Circuit Diagram ...
Page 71: ...4000 Evaluation Unit Service Manual 4 39 4 17 2 Circuit Diagram ...
Page 73: ...4000 Evaluation Unit Service Manual 4 41 4 18 2 Circuit Diagram ...
Page 75: ...4000 Evaluation Unit Service Manual 4 43 4 19 2 Circuit Diagram ...
Page 77: ...4000 Evaluation Unit Service Manual 4 45 4 20 2 Circuit Diagram ...
Page 79: ...4000 Evaluation Unit Service Manual 4 47 4 21 2 Circuit Diagram ...
Page 124: ...4 5 Gas Measuring Instruments Unit AVL 4000 5 28 Service Manual ...
Page 149: ...4000 Opacity Measuring Chamber Type A Service Manual 6 25 Fig 6 30 Fig 6 31 ...
Page 150: ...Opacity Measuring Chamber Type A AVL 4000 6 26 Service Manual Fig 6 32 ...
Page 180: ...Spare Parts Lists AVL 4000 8 14 Service Manual ...
Page 182: ...Tools for the Service Technician AVL 4000 9 2 Service Manual ...