4000
4-/5-Gas Measuring Instruments Unit
Service Manual
5-15
Primary error symptom Primary cause
(of error)
Secondary cause of
error
Possible tests/
establishing cause
Remedy
Communication
error
RS232 cable, Andros
measuring
instruments unit
faulty
Cable connected
incorrectly
Service mode:
pumps and valves
can be switched but
no measurement
values. Check cable
connections
Connect cable
correctly or replace
measuring
instruments unit
Zeroing in service
mode
Not an error; can
take up to 80 sec
Measurement cell
faulty/soiled
Condensate end of
pump is faulty.
Condensate
discharge tube is
clamped closed
Raw values for CO,
H, CO
2
lower than
80000.
Check hoses, test
flow rate
(faster than 2
min
l
per channel)
Clean or replace
pump
Clean or replace
measurement cell
Replace measuring
instruments unit
Faulty light source Light source faulty,
power supply unit
faulty
UC raw values in
service mode lower
than 10000
ES not zero
Light source does
not light up
Replace light source
Replace measuring
instruments unit
No measurement
values displayed but
chopper is working
Faulty detector
UC raw values in
service mode lower
than 10000
ES not zero
Replace detector
Replace measuring
instruments unit
Faulty transformer
Faulty fuse
Power supply,
pneumatics board or
pump
Normal:
resistance of
transformer
windings approx.
50
Ω
, very high
resistance to ground
Replace transformer
(465)
Replace fuse
Faulty power
supply
Disconnect power
supply, measure
transformer
resistance
Change power
supply
Faulty pneumatics
board (465)
Faulty driver on
board
Driver on
pneumatics board/
cooling element is
very hot
Replace measuring
instruments unit
Replace board
Replace transformer
too if necessary
Mains short circuit
Faulty pump
Short circuit in
pump electronics
(465)
Disconnect pump,
measure
transformer
resistance (approx.
50
Ω
normal),
ground fault?
Replace pump
Summary of Contents for AVL DiCom 4000
Page 4: ......
Page 6: ......
Page 22: ...Periodic Maintenance AVL 4000 1 12 Service Manual ...
Page 47: ...4000 Evaluation Unit Service Manual 4 15 4 9 Main Board 4 9 1 Components Location Diagram ...
Page 48: ...Evaluation Unit AVL 4000 4 16 Service Manual 4 9 2 Circuit Diagrams ...
Page 49: ...4000 Evaluation Unit Service Manual 4 17 ...
Page 50: ...Evaluation Unit AVL 4000 4 18 Service Manual ...
Page 51: ...4000 Evaluation Unit Service Manual 4 19 ...
Page 52: ...Evaluation Unit AVL 4000 4 20 Service Manual ...
Page 54: ...Evaluation Unit AVL 4000 4 22 Service Manual 4 10 2 Circuit Diagrams ...
Page 55: ...4000 Evaluation Unit Service Manual 4 23 ...
Page 56: ...Evaluation Unit AVL 4000 4 24 Service Manual ...
Page 57: ...4000 Evaluation Unit Service Manual 4 25 ...
Page 59: ...4000 Evaluation Unit Service Manual 4 27 4 11 2 Circuit Diagram ...
Page 61: ...4000 Evaluation Unit Service Manual 4 29 4 12 2 Circuit Diagram ...
Page 63: ...4000 Evaluation Unit Service Manual 4 31 4 13 2 Circuit Diagram ...
Page 65: ...4000 Evaluation Unit Service Manual 4 33 4 14 2 Circuit Diagram ...
Page 67: ...4000 Evaluation Unit Service Manual 4 35 4 15 2 Circuit Diagram ...
Page 69: ...4000 Evaluation Unit Service Manual 4 37 4 16 2 Circuit Diagram ...
Page 71: ...4000 Evaluation Unit Service Manual 4 39 4 17 2 Circuit Diagram ...
Page 73: ...4000 Evaluation Unit Service Manual 4 41 4 18 2 Circuit Diagram ...
Page 75: ...4000 Evaluation Unit Service Manual 4 43 4 19 2 Circuit Diagram ...
Page 77: ...4000 Evaluation Unit Service Manual 4 45 4 20 2 Circuit Diagram ...
Page 79: ...4000 Evaluation Unit Service Manual 4 47 4 21 2 Circuit Diagram ...
Page 124: ...4 5 Gas Measuring Instruments Unit AVL 4000 5 28 Service Manual ...
Page 149: ...4000 Opacity Measuring Chamber Type A Service Manual 6 25 Fig 6 30 Fig 6 31 ...
Page 150: ...Opacity Measuring Chamber Type A AVL 4000 6 26 Service Manual Fig 6 32 ...
Page 180: ...Spare Parts Lists AVL 4000 8 14 Service Manual ...
Page 182: ...Tools for the Service Technician AVL 4000 9 2 Service Manual ...