ZM400 Series Indicators Service Manual
77
6.5
Ports
l
AC Output 4 relays (20-240VAC) card
l
DC Output 4 relays (3-60VDC) card
Depending on the choice made for
tYPE
, a
conFig
menu may appear. See the Options
menu on page 76. See the Options menu on page
.
If USB-D or RS485/20ma card is installed:
Config
bAud
Select from the list of baud rates.
d-bitS
Select 7 or 8.
PAritY
Select none, odd or even.
S-bitS
Select 1 or 2 stop bits.
The Serial RS485/20ma and USB-D options require a Bind setting of
oPtion
If Analog card is installed:
Config
LoW
Set the displayed value for the lowest analog
output.
Hi
Set the displayed value for the maximum analog
output.
cALib
ZEro
Set the analog output zero point.
SPAn
Set the analog output span point.
AdJuSt
Adjustments are normally necessary to set the
zero or span output values to correlate with the
connected analog device. Pressing the following
keys will adjust the signal output as noted.
TARE
increase in small increments
SELECT
decrease in small increments
increase in large increments
UNITS
decrease in large increments
Press
ZERO
to complete the analog output calibration
The Analog Output requires a Bind setting of
oPtion