40
ZM400 Series Indicators Service Manual
6
ADMIN level menus
6.2
Calibration Procedure
Use the Calib menu to perform Zero and Span calibration, add Linearity correction
points, manually input calibration parameters for Zero and Span, manually input
Gravitational correction values, view the live weight, set the calibration unit of measure
and print out a calibration report. Follow the menu in
and the steps that
follow.
Figure 6.3 Calibrate menu
Access the calibration procedure directly using the calibration password, 2580, or
access it through the Setup menu, password 3088. See
6.2.1 Scale 1-2
Select the scale to be calibrated, Scale 1 or Scale 2. The
must be set to 2 to access Scale 2 settings.
Calib
Zero
Linear
Input
Gravity
Last
Cal.Zero
Temp
Zero
Span
Counts
mV
Counts
mV
XXXX
USB
Port 1
Port 2
1000g
Cal.Unit
LB
Display
G-fact
Lat
Alt
Calc
Pt2
Pt9
Span
Scale 1
Scale 2
View or
enter new
span
Live weight
displayed.
Place span
weight on
scale
Enter
Pt 2 test
weight
Live weight
displayed.
Remove
all weight.
Live weight
displayed.
Live weight
displayed.
Place Pt 9
weight on
scale.
Enter
Pt 9 test
weight
View or
enter
G-factor
Enter
latitude
View
live weight
(F1 to stop)
Enter
altitude
Calculate
s gravity
factor
Live weight
displayed.
Live weight
displayed.
Live weight
displayed.
Place Pt 2
weight on
scale.
Live weight
displayed.
to