ZM400 Series Indicators Service Manual
53
6.3
Scale
M.intrVL
Multi-interval - the division size will change immediately as it enters a new
weight range.
6.3.10 Type
tYPE
Use this to select the mode and corresponding binding for the selected
scale.
ModE
Select the mode from these choices:
oPtion
Pick this mode if you are using the analog scale
option card.
ProtcL
Pick this mode if you are using the BSQ base
connected to a serial port.
diSAbLE
Pick this to disable the selected scale. This is the
default setting for Scale 2.
onbrd
Pick this if you are using the onboard analog
scale. This is the default setting for Scale 1.
bind
This option appears in two different forms depending on which
Mode you pick. If you pick
diSAbLE
or
onboArd
in the Mode
menu, the
bind
menu does not appear.
If you choose
oPtion
, this selection requires a bus and card
number. For the ZM400 series indicator, only one option card
can be installed so these values are already defaulted to bus
1 and card 1 and do not need to be configured.
If you choose
ProtcL
, you must choose a binding with a
protocol selection. There are 10 available protocols and the
protocol number selected must be matched with a properly
configured protocol shown in
This completes the Scale menu.
Multi-interval is not an agency approved mode of operation.
Type
Mode
Option
Protcl
Disable
Onbrd
Bind
Bus
Card
Bus X
Card X
Bind
Prot 1
Prot 10
to
If Mode is
Option
If Mode is
Protcl