Error 37
Unknow type of Digi probe.
Please update firmware version and after that detect Digi
probe(s) again.
Error 38
Communication error with memory for calibration data inside
Digi probe. It is likely that Digi probe is not connected properly,
or probe is damaged.
Please connect properly Digi probe or replace damaged
probe.
Error 39
Memory for calibration data inside Digi probe have wrong
CRC.
Please follow procedure for Error 36.
Error 40
Connected type of Digi probe is not same as was detected
probe.
Please detect connected Digi probe(s) again.
Error 50
Device configuration for channel is damaged.
Damaged configuration can be fixed by the Factory defaults
procedure or using WifiSensorUtility which can be obtained
from technical support.
Error 52
Measured value cannot be shown due to overflow during
conversion.
Error 53 or n/a
Value is not available. This error is shown at disabled channels
or when value was not measured yet. CO
2
concentration is
available 15 sec after device start-up.
Error 55
This error is related to values transferred via Modbus TCP. It
signalises overflow of Modbus register.
Warning exclamation mark on LCD
Warning exclamation mark shown on left bottom corner of LCD display
signalise issue with device itself. Main reasons for this warning are:
•
Hardware issue with one of these components
– WiFi coprocessor,
internal memory chip, RTC chip or LCD driver chip.
•
Damaged content of configuration memory.
Damaged content of configuration memory can be identified by warning
message in case of attempt to device setup. Damaged content can be fixed
by the
or using WifiSensorUtility. This software
can be obtained from technical support.
In case of issue with hardware, please contact technical support directly.