Settings alarm for a channel
Each channel allows for configuration of two separate alarms. Alarm mode
selects direction of alarm
– Lower than limit, Higher than limit, or Disabled.
Limit value is a threshold for activation of alarm. Alarm is activated when
measured value exceed selected limit for a quantum of time set by Delay
option. Alarm state is cleared when measured value returns under set limit
with selected Hysteresis. When alarm is activated it can be acoustic or optical
LED signalisation activated. Alarm on channel can be activated optionally
when measurement error on channel occurred.
WiFi sensors have capability to send email in case of alarm on channel is
activated. Up to four email recipients can be configured. Addresses of email
recipients are configurable at
. Each recipient can be